问个microtiter plate reader样品稀释的问题# Biology - 生物学
h*r
1
我圡,没怎么用过microtiter plate reader,都是用分光光度计来着,但是太费比色
皿。准备去别的实验室蹭microtiter plate reader,问题是OD超过1或者2是不是就要
稀释呢?问过他们实验室的人,也不清楚。请教了!
另外用nanodrop测DNA呢?上次测了个500ng/ul,应该也就不准了吧?
多谢!
avatar
s*g
2
啥试验啊?
avatar
i*0
3
To be honest, according to my experiences, the worst reliability of a plate
reader is the absorption assays. I have tried Bradford or OD280, both give
high standard deviation. However, if you just want a guess, try using your
sample at original conc, 1:1 dilution, and 1:4 dilution in triplicate. You'd
better have BSA standards at least 5 concentrations in the same plate in
triplicate. These will give you a relatively reliable readout. OD is linear
at 0.5 to 1, roughly. good luck
avatar
i*0
4
If you just want to normalize protein concentration for immunoblot, just use
a nano drop and read OD280.
avatar
h*r
5
我做微生物的lacz assay,手上待测和重新engineer的promoter有30多个吧,用lacZ的
酶活来看不同promoter的activity.有的还得用不同的诱导条件。
想在自己实验室反应完全后再出去测。
没想到这么挫啊!用分光光度计测还行,就是累和太费比色皿。